BE ORIGINATIVE NICHIRYO

ISO9001ISO14001

  • Company Profile
  • Environment
  • Japanese
  • Nichiryo America,Inc

NSP-7000R Multi-Channel Auto Sampling SystemPDF

Smooth operation Come back as more silent system Small foot-print design Compact design for desk top operation Simple keypad operation No PC requirement

NSP-7000R is the compact liquid sampling system for desk top operation.
NSP-7000R is more silent and smooth than past best seller model NSP-7000.
This is the perfect system for microplate operation such as reagent dispensing, sample dispensing, serial dilution and on-plate transferring.

FLEXIBLE SYSTEM

  • The new disposable tips are replaced automatically to ensure no carry-over operation.
  • High speed and safety operation with good accuracy and precision.
  • 3 positions sliding work stage accepts disposable tip rack, reagent reservoir and microplate in 8ch or 12ch way.
  • Compact design for desk top operation.
  • No PC requirement, the built-in keypad panel is designed for easy programming and operation.
    There are five variable dispensing modes.
     Mode 2 @Reagent dispensing
     The reagent can be transferred to a labware in a row of wells from a vessel for one time dispensing.
     Mode 3 @Continuous dispensing of reagents and samples.
     The reagent can be transferred to a labware in multiple wells for sequential dispensing.
     Mode 4 @Serial dilution
     Perform serial dilution in almost any ratio.
    Perform serial dilutions either 8 wells across or 12 well across.
     Mode 5 @Plate to plate transfer of samples
     Transfer samples to or from microplates to an assay plate.
    Perform replica plating.
     Mode 6 @On-plate transfer (moving within the same plate)
     Pick-up and dispensing in a row one after another on the same labware at the mixing stage
  • Link files for complex liquid handling requirement by Mode 7. (maximum 100 files)
  • Easy to call the saved protocol to operate from memory.
  • Wide volume range of 10 - 220ul in increments of1ul.
  • Perfect for examination of HIV and HCV by condensation method and ELIZA method.
  • Microplate mixer installed as standard option.
¡Technical Data
Dimension
Dimension to Install
300mm(W) x 645mm(D) x 585mm (H)
 300mm(W) x 810mm(D) x 585mm (H)
WeightApprox. 42kg
Usable tipsNICHIRYO HT-FSR for robotic system irackedj
Usable microplatesSBS standard 96 well or 384 well.
( for deep-well plate or mini-tube plate, please contact us)
Nozzle head220ul x 12ch
Sampling volume
(warranty range)
(Setting is available
from 1ul)
10 - 220ul
Accuracy and CV<}2.8% AC, <2.8% CV at 10ul
<}0.8% AC, <0.8% CV at 220ul
Working stage3 positions FTip rack, Reservoir (30mlx4 containers), Microplate
Liquid surface sensingN/A
Dispensing speed1 microplate dispensing 100ul with 12ch head - approx. 45 sec.
Microplate mixerBuilt-in working stage
Power sourceAC 100 - 120V or AC 200 - 240V 50/60Hz 100VA (100V)

SPECIFICATION AND OPTIONS

¡System
Specification
00-NSP-7000RSliding work stage, 220ul x 12ch nozzle head, plate mixer
¡HT series tips for robotic system and reservoir
No.NameVolumeContents
00-HT-FSBulk tip for robotic system10 - 220ul1000
00-HT-FSRRacked tip for robotic system10 - 220ul960
00-HT-FSR-BRacked tip for robotic system
(sterilized)
10 - 220ul960
00-NSP-RSV4 tray reservoir-1
00-NSP-RSV-NFTray for reservoir-4
00-NSP-RSV80Large volume reservoir
(without frame)
-50

–Specification is subject to change without notice.
–WINDOWS is the trademark of Microsoft company.
–Pentium is the trademark of Intel company.

Products TOP@Page TOP